LC-MS
European and US regulators now allow less-than-lifetime adjustments to nitrosamine acceptable intakes. For quality control laboratories, every limit-based decision threshold moves with them.
Nitrosamine testing has always been set by one number: the acceptable intake (AI) for each N-nitroso impurity. That number decides how low a confirmatory method must quantify. This autumn, the European Medicines Agency (EMA) and the US Food and Drug Administration (FDA) both changed how it is calculated. The change affects medicines taken for less than a lifetime.
On Thursday, 24 September 2026, the FDA’s Center for Drug Evaluation and Research (CDER) updated its nitrosamine acceptable intake limits page under docket FDA-2020-D-1530. A new section sets out how less-than-lifetime (LTL) adjustment factors apply to nitrosamine impurities.
The FDA says this is consistent with the LTL framework in ICH M7(R2), the International Council for Harmonisation (ICH) guideline on mutagenic impurities. Its table multiplies the recommended lifetime AI by 80 for treatment of one month or less. The factor is 13.3 for more than one month up to 12 months, and 6.67 for more than one year up to ten years. Treatment beyond ten years takes the lifetime AI unchanged.
These factors mirror the duration-based intakes in Table 2 of ICH M7. That table allows 120 µg/day for up to one month, 20 µg/day up to 12 months and 10 µg/day up to ten years, against 1.5 µg/day for lifetime exposure.
The FDA attaches conditions. Treatment duration must match the labelled indication and dosing regimen. Where a label covers both short-term and long-term use, the agency generally calculates on long-term use. For intermittent dosing, it counts total dosing days across a lifetime. Where several nitrosamines are present, total exposure should not exceed a cancer risk of one in 100,000. LTL adjustment is not recommended for interim AI limits.
In Europe, the EMA updated its guidance for marketing authorisation holders on Thursday, 5 November 2026. Its Committee for Medicinal Products for Human Use (CHMP) agreed a revised question-and-answer document, EMA/409815/2020 Rev. 24, in alignment with the Coordination Group for Mutual Recognition and Decentralised Procedures – Human (CMDh).
According to the EMA’s summary, the new question 24 allows higher AI limits by applying the ICH M7 LTL approach to eligible medicines used for less than ten years. The exact factors and eligibility conditions in the Rev. 24 text could not be checked for this article.
This is a reversal of the previous position. Rev. 23, dated Friday, 10 October 2025, stated that the LTL approach “should not be applied” when calculating limits. It allowed LTL only as a temporary measure after consultation with competent authorities.
Other changes listed by the EMA include a CAPA (corrective and preventive action) period extendable from three to seven years under certain conditions. There is also a new risk factor for parenteral products in polymer bags. Further changes cover other N-nitroso compounds and limits for nitrosamines arising from excipients. Appendix 1, the EMA’s list of established AIs (EMA/42261/2025 Rev.14), was most recently updated on Tuesday, 13 October 2026.
Neither regulator has set a new sensitivity requirement. The thresholds that govern method sensitivity are expressed relative to the AI-based limit, so they shift whenever the limit does.
In Rev. 23, the EMA set three tiers. For routine quantitative control, the limit of quantification (LOQ) should be at or below the acceptable limit. For skip testing, it should be no more than 30 per cent of that limit. To omit a specification, it should be no more than 10 per cent.
The FDA’s September 2024 guidance on controlling nitrosamine impurities says LOQs should be “commensurate with the level at which the impurities must be controlled”. Its LTL section states that a specification is generally not needed where levels are consistently no more than 10 per cent of the recommended AI limit.
The FDA converts an AI to a concentration by dividing the AI in ng/day by the maximum daily dose (MDD) in mg, giving parts per million (ppm). As an illustration, take 26.5 ng/day, the AI the FDA recommends when no compound-specific limit can be determined, and an MDD of 500 mg. The lifetime limit is 0.053 ppm. With the FDA’s 80-fold factor for treatment of one month or less, it becomes 4.24 ppm. With the 13.3-fold factor, it is around 0.70 ppm. These figures are this article’s own calculation, not regulatory values.
A method validated against a lifetime limit is not made unfit by a higher limit. The changes lie in the arithmetic around it.
The first point is the denominator. Specification limits, skip-testing thresholds and the 10 per cent cut-off for omitting a specification all depend on which AI applies. Laboratories applying an LTL-adjusted AI will need to recalculate each of them from that figure.
The second is the duration basis. The FDA’s conditions on labelled duration, mixed short-term and long-term use, and intermittent dosing determine which factor applies. A method set up for one product may carry assumptions that do not hold for another containing the same impurity.
The third is multiple impurities. The FDA’s cumulative risk condition means individual limits cannot simply be scaled in isolation. Methods covering several nitrosamines may need reporting that supports a combined assessment.
The fourth is high-dose products. Rev. 23 anticipated exceptions for medicines used at high daily doses, where ppm limits fall lowest. For short-duration high-dose products, a higher AI raises the ppm target that trace-level methods must reach.
On technique, Rev. 23 required accurate mass approaches: tandem mass spectrometry (MS/MS), as in liquid chromatography–tandem mass spectrometry (LC-MS/MS), or high-resolution accurate mass systems. The FDA’s LTL section asks for “sensitive and appropriately validated methods”. Its 2024 guidance recommends confirmatory testing on at least three representative batches.
Whether Rev. 24 alters the EMA’s question 9 LOQ wording has not been confirmed for this article. Laboratories working to EMA requirements will need to read the revised document itself.