• Total Nitrosamines Analysis System

Autosamplers

Total Nitrosamines Analysis System

Feb 18 2013

In certain production processes such as those in cosmetics or agricultural pesticides industries, it is important to analyse for the presence of Nitrosamines, but it is not all ways essential to know the exact profile of the nitrosamine species.

The Ellutia Chemical Stripping System has been developed to be used in unison with an 800 Series TEA to perform total nitrosamine analysis. Instead of using a pyrolyser to heat the nitrosamine to remove the NO group, a chemical reaction can be employed. The alleviation of NO from the nitrosamine is performed by a reflux reaction with hydrobromic acid in ethyl acetate. The nitrosamine sample is injected into the reaction vessel and produces NO, a secondary amine and bromine. The NO is then carried in a gas stream through a cold trap (removing any vapour) to the TEA

For further information please visit

http://www.ellutia.com/TotalNitrosamines.html
 


Digital Edition

Chromatography Today - Buyers' Guide 2022

October 2023

In This Edition Modern & Practical Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...

View all digital editions

Events

HUPO 2024

Oct 20 2024 Dresden, Germany

SETAC North America

Oct 20 2024 Fort Worth, TX, USA

HPLC

Oct 21 2024 Dalian, China

Intech

Nov 12 2024 Tel Aviv, Israel

analytica China

Nov 18 2024 Shanghai, China

View all events