Autosamplers
Analysis of Total Nitrosamines
Nov 14 2014
In certain production processes such as those in cosmetics or agricultural pesticides industries, it is important to analyse for the presence of Nitrosamines, but it is not always essential to know the exact profile of the nitrosamine species.
The Ellutia Chemical Stripping System has been developed to be used in unison with an 800 Series TEA to perform total nitrosamine analysis. Instead of using a pyrolyser to heat the nitrosamine to remove the NO group, a chemical reaction can be employed. The alleviation of NO from the nitrosamine is performed by a reflux reaction with hydrobromic acid in ethyl acetate. The nitrosamine sample is injected into the reaction vessel and produces NO, a secondary amine and bromine. The NO is then carried in a gas stream through a cold trap (removing any vapour) to the TEA.
For further information please click here.
Digital Edition
Chromatography Today - Buyers' Guide 2022
December 2021
In This Edition Modern & Practocal Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...
View all digital editions
Events
Jun 04 2023 Houston, TX, USA
Plastics Recycling World Expo Europe
Jun 14 2023 Essen, Germany
Jun 18 2023 Dusseldorf, Germany
Jul 04 2023 Cardiff, UK
Jul 11 2023 Lagos, Nigeria