Analysis of oxidised monoclonal antibodies using YMC's BioPro HIC BF

Sep 22 2020 Read 487 Times

During the manufacture and/or storage of biopharmaceuticals, variants with different properties from desired substances are produced by enzyme reactions or physicochemical interactions. Characterisation of these variants is of great importance from the perspective of ensuring efficacy and safety of pharmaceutical products.

Oxidised variants of monoclonal antibodies (MAbs) can be analysed by hydrophobic interaction chromatography (HIC). In this application note, the separation of MAb samples and their oxidised species using YMC’s HIC column, BioPro HIC BF, is described. BioPro HIC BF is a more hydrophobic HIC phases dedicated for proteins of low hydrophobicity and oxidised MAb variants.

MAb oxidation with t-BHP treatment

tert-Butyl hydroperoxide (t-BHP) was used as a chemical oxidant to promote oxidation of methionine residues of NIST MAb. Subsequently, papain was used for the preparation of two Fab fragments and one Fc fragment.

Analysis of oxidised MAbs

When analysing the oxidised NIST MAb using the ammonium sulphate (a), four peaks appeared at earlier elution times compared to the peak of the non-oxidised MAb, presumably due to the conformational changes via the oxidisation of the methionine residues.

Applying sodium chloride (b), eight peaks appeared at earlier elution times compared to the peak of the non-oxidised MAb. The better resolution was further achieved within shorter analysis time compared to the usage of ammonium sulphate (a).

Analysis of papain-digested oxidised MAb

The papain digests of NIST MAb samples with and without oxidisation were analysed. The Fab and Fc fragments were characterised from the chromatogram of the papain digested MAb.

In the chromatogram of the papain digested oxidised MAb, multiple peaks appeared at earlier elution times compared to the peaks assigned to the Fab and Fc fragments. These peaks correspond to oxidised fragments according to a previous report [1], describing that oxidised fragments elute earlier than the non-oxidised ones.

[1] Journal of Chromatography A, 2008, 1214, 81-89

Reader comments

Do you like or dislike what you have read? Why not post a comment to tell others / the manufacturer and our Editor what you think. To leave comments please complete the form below. Providing the content is approved, your comment will be on screen in less than 24 hours. Leaving comments on product information and articles can assist with future editorial and article content. Post questions, thoughts or simply whether you like the content.

Post a Comment





Digital Edition

Chromatography Today - September 2020

September 2020

In This Edition Articles Fundamental Aspects - Analysis and Identification of Mezcal and Tequila Aromas by Ambient Ionisation MS, GC-MS, and GCxGC-MS - Analysis of Polynuclear Aromatic Hy...

View all digital editions

Events

Eastern Analytical Symposium

Nov 16 2020 ONLINE EVENT

IMSC 2021 - NEW DATES

Feb 07 2021 Rio de Janeiro, Brazil

PITTCON 2021

Mar 06 2021 New Orleans, LA, USA

ACS National Meeting & Expo, Spring 2021

Mar 21 2021 Philadelphia, PA, USA

Arabchem

Mar 22 2021 Dubai, UAE

View all events