5 Gas Management Tips for GC

Jul 30 2014

Gas management is important for gas chromatography systems (GC) for both analytical and practical considerations. Looking after the gas supply helps to ensure accurate results, and to reduce costs and downtime.

Gas Cylinders

Never allow gas cylinders to run dry. As the gas pressure in the cylinder drops, the chance that contaminants (like water or hydrocarbons) can desorb from the cylinder increases. These contaminants can then enter the gas supply. The contaminants can have a negative effect on the system, leading to increased costs and downtime as cleaning and flushing takes place.

It is recommended that gas cylinders be fitted with a low level alarm. This will allow an operator to switch over gas cylinders before contamination can occur. Values used for setting the alarm include:

  • a fixed value of 150 psi is sometimes used, or
  • a value of 10% of the initial cylinder pressure can be used. Sometimes suppliers only guarantee the cleanliness of the supply down to this value.

Purifiers

Installing gas purifiers in the system reduces the risk of contaminants entering the column and the detector, especially if high purity gases are not used. Purifiers can increase column life and sensitivity. Various types of purifiers are available, including combination or single function. It is important to place single function filters in the correct order; from the gas supply this is:

  • moisture filter
  • hydrocarbon filter, then
  • oxygen filter.

Stainless steel and glass are considered the best materials to use for the bodies of purifiers. It is important that purifiers are installed correctly. If the filter should be vertical, it must be installed vertically, otherwise gas can bypass the filter material. Ensure all the equipment is of chromatography standard, and has been cleaned before use.

Tubing and Lines

For a high purity gas supply, it recommended that only stainless steel or copper tubing is used. Teflon, nylon and other types of plastic tubing can contain contaminants that reduce gas purity. Also some plastics are permeable to oxygen and moisture, and these can contaminate the gas supply.

Location

Keep the tubing length between the gas cylinders and the GC as short as possible. Also reduce the number of connections to a minimum. The possibility of a gas leak increases with the number of connections in the gas supply line. Leaks cost money and can be a safety issue as some GC gases are flammable. It is important to check for gas leaks regularly, and after changing purifiers or tubing sections. A leak detector can be fitted to the system to automatically check for leaks.

Regulators

Make sure that the correct regulator is chosen for each cylinder: different gases require different regulators. It is important to purge the air from new regulators before use, as the air can damage the GC system.

Regulators should be set at a pressure 20 psi higher than the pressure required in the GC method, especially when running methods with high pressures between 100-150 psi.

Hydrogen

Due to the decreasing levels of helium, numerous laboratories are looking for alternative carrier gases. For many GC applications, hydrogen provides a suitable, if not better, substitute. This article, How Using Hydrogen Carrier Gas Can Alleviate Your Helium Supply Woes, explains this in more detail. 

Stock Image


Digital Edition

Chromatography Today - Buyers' Guide 2022

October 2023

In This Edition Modern & Practical Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...

View all digital editions

Events

Korea Lab 2024

Apr 23 2024 Kintex, South Korea

Korea Chem 2024

Apr 23 2024 Seoul, South Korea

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

SETAC Europe

May 05 2024 Seville, Spain

ChemUK 2024

May 15 2024 Birmingham, UK

View all events