Determination of Nimesulide and Thiocolchicoside Impurities in a Topical Gel Product UPC2

Aug 20 2014

Author: Gülçin Tok, Yelda Erdem, Ediz Yıldırım, Ali Türkyılmaz, Ümit Çifter on behalf of Sanovel Pharmaceuticals

Free to read

This article has been unlocked and is ready to read.

Download

A topical pharmaceutical product consisting of Thiocolchicoside and Nimesulide was analysed using UPC2 for determination of impurity content.  Historically, these impurities were analysed by normal phase and reversed phase LC, especially in the case of the Thiocolchicoside impurities. UPC2 offered a single platform technique instead of these two. For each drug substance, separate methods were developed and validated with reference to pharmacopoeial requirements. The sample preparation for both UPC2 methods is identical and is significantly simpler than the liquid chromatographic methods which included a solid phase extraction step.

Introduction

Muscle relaxants are used alone or together with nonsteroidal anti-inflammatory drugs (NSAIDs) in treating pain [1].  Thiocolchicoside (TH) belongs to the muscle relaxant class of medications and reduces muscle spasms while Nimesulide (NIM) is an NSAID that reduces pain and inflammation. TH has never been used together with NIM in combination as a topical gel formulation for treating inflammatory pain and muscle skeletal system diseases.  A literature survey revealed that several analytical methods have been described for analysis of TH as single component or in combinations with other drugs [2]. Hence, no pharmacopoeial methods or any other studies include TH and NIM together in a combination formulation. 

The main challenges to the analysis of this product were the high number of impurities requiring separation, and the low amount of TH and NIM in the drug product requiring significant sample preparation prior to analysis.  The sample concentration needed to consider the maximum daily dose and the quantification limits (LOQ) of TH and NIM, and therefore be high enough to obtain a detectable response.  A high sample concentration may also be problematic for LC systems as it may result in a pressure increase and may also be detrimental for column lifetime.

The present study demonstrates the determination of all TH and NIM impurities using two separate validated UPC2 methods with high efficiency and resolution, shorter separation times, reduced solvent consumption and sampling time, streamlined sample preparation, and lower cost of analysis per sample when compared to the LC methods. 

UPC2 is a powerful tool and is often described as having GC like efficiency, but with wide molecular applicability.  It can reduce sample preparation and analysis time by direct injection of organic solvents/extracts and is an orthogonal separation technique to LC which means it can provide higher confidence in identifying impurities/degradants, and it can allow full sample characterisation and separation of analytes from matrix interferences without the need for complex sample preparation [3].

 

Experimental

Thiocolchicoside and its related substances were analysed using the following conditions. Columns: ACQUITY UPC2 BEH (2.1 x 100 mm, 1.7 µm) column and ACQUITY UPC2 BEH (2.1 x 5 mm, 1.7 µm) VanGuard pre-column were used at 35ºC with gradient elution at a flow rate of 1.3 mL/min; UV detection at 258 nm; the ABPR (Active back pressure regulator) was set to 1800 psi; the injection volume was 7.5 µL and sample tray temperature was thermostated at 8ºC.  The carbon dioxide mobile phase was modified with a co-solvent consisted of 0.01 M ammonium acetate in methanol, 2-propanol, ammonia (25 %w/v) (920:80:12, v/v/v).  The LC gradient program was set as (time (min)/% co-solvent): 0/1.0, 5/18.0, 15/20.0, 17/1.0 and 20/1.0. 

Nimesulide and its related substances were analysed using the following conditions.  Column: ACQUITY UPC2 BEH-2EP (2.1 x 100 mm, 1.7 µm) column and ACQUITY UPC2 BEH-2EP (2.1 x 5 mm, 1.7 µm) VanGuard pre-column were used at 48ºC with gradient elution at a flow rate 1.0 mL/min; UV detection at 230 nm; the ABPR was set to 1500 psi; injection volume 2 µL; and sample tray temperature at 8ºC. The carbon dioxide mobile phase was modified with a co-solvent consisted of methanol and absolute ethanol (1:1 v/v). The gradient program was set as (time (min)/% co-solvent): 0/0.1, 3/1.0, 10/8.0, 12/8.0, 14/28.0, 17/28.0, 18/0.1 and 20/0.1. 

The sample solutions for both methods were prepared as follows. Approximately 10.0 g of topical gel (equivalent to 100 mg NIM and 25 mg TH) weighed into a 100 mL volumetric flask. 50 mL acetone was added and shaken continuously to dissolve. This solution was made to volume with acetone and sonicated for 10 minutes. The solvent was then evaporated from the solution under 300-350 mbar vacuum and the solvent free solution transferred into a 20 mL volumetric flask and made to volume with acetone. (CTH:1250 µg/mL, CNIM:5000 µg/mL)

Spiked test solution for TH was prepared by transferring 10.0 g topical gel into a 100 mL volumetric flask, 50 mL of acetone added and shaken to dissolve and made to volume with acetone. This solution was then sonicated for 10 min. The solvent was then evaporated fromthe solution under 300-350 mbar vacuum and the solvent free solution transferred into a 20 mL volumetric flask. 0.125 mg of all known TH impurities were added to the same volumetric flask and made to volume with acetone. (CTH:1250 µg/mL, CNIM:5000 µg/mL, CColchicine:6.25 µg/mL, CNDNFT:6.25 µg/mL, CCholchicoside:6.25 µg/mL)

Spiked test solution for NIM was prepared by transferring 10.0 g topical gel into a 100 mL volumetric flask, 50 mL of acetone added and shaken to dissolve and made to volume with acetone. This solution was then sonicated for 10 min. The solvent was then evaporated from the solution under 300-350 mbar vacuum and the solvent free solution transferred into a 20 mL volumetric flask. 0.150 mg of all known NIM impurities were added to the same volumetric flask and made to volume with acetone. (CTH:1250 µg/mL, CNIM:5000 µg/mL,  CNIM imp A:7.50 µg/mL, C NIM imp B:7.50 µg/mL, C NIM imp C:7.50 µg/mL, C NIM imp D:7.50 µg/mL, C NIM imp E:7.50 µg/mL)

A placebo solution was prepared by transferring approximately 10.0 g topical gel placebo into a 100 mL volumetric flask, 50 mL of acetone added and shaken to dissolve and made to volume with acetone. This solution was then sonicated for 10 min. The solvent from the solution was then evaporated under 300-350 mbar vacuum and the solvent free solution transferred into a 20 mL volumetric flask, again making to volume with acetone. 

All solutions were filtered through 0.2 µm GHP PALL, PTFE hydrophilic or PTFE hydrophobic filters.

 

Results and Discussion

The method parameters given in the Experimental section were determined depending on the chemical properties of the drug substances such as their solubility, polarity, log P and pKa values. Using UPC2 as the separation system, no high pressure issues were observed with the high concentration sample (0.5 g topical gel/1 mL diluent). The same sample preparation was used for both TH and NIM methods. Preparation of the sample solution was uncomplicated, solvent friendly and less time-consuming than the comparative LC method. Although solid phase extraction was employed with the original assay, which used a LC separation mode, the column lifetime was quite short (less than 10 injections) and repeated injections could not be obtained. Three different methods were used in the LC methods which included a reverse phase for NIM, a normal phase and a reverse phase for TH. Normal phase LC methods for TH used the following conditions. Column: Silica (4.6 x 250 mm, 5 µm) at 25ºC; isocratic flow rate 1.5 mL/min; UV detection at 360 nm; injection volume 20 µL; mobile phase consisted of heptane, chloroform, methanol, acetic acid (35:5:11:1, v/v/v/v); analysis time was 60 min. Reverse phase methods for TH used the following conditions. Column: C18 (4.6 x 250 mm, 5 µm) at 25oC; isocratic flow rate 1.5 mL/min; UV detection at 290 nm; injection volume 20 µL; mobile phase consisted of mixture of 45 volumes of methanol and 55 volumes of a 2.8 g/L solution of ammonium dihydrogen phosphate adjusted to pH 7.3 with tri ethyl amine; analysis time was 60 min. LC methods for NIM used the following conditions. Column: ODS (3.9 x 150 mm , 4.0 µm) at 30ºC; isocrotic flow rate 1.3 mL/min.; UV detection at 230 nm; injection volume 20 µL; mobile phase consisted of a mixture of 35 volumes of acetonitrile and 65 volumes of a 1.15 g/L solution of ammonium dihydrogen phosphate adjusted to pH 7.0 with ammonia; analysis time was 60 min. The advantages of UPC2methods over the comparative HPLC methods were shown in the Table 1. 

Acetone was chosen as the diluent since both TH and NIM were soluble in it, miscible with water coming from the gel product and also compatibility with both CO2 and the co-solvents selected.  In the TH method, better resolution was obtained between TH and N-Deacetyl N-formyl thiocholchicoside (NDNFT) when the percentage of ammonia in the co-solvent was 1.2 %v/v. Although the resolution was increased by increasing the ammonia content, baseline noise became worse. As TH is more polar than NIM, it was more retained on column, and its retention time was correspondingly later than that of NIM.

The elution program of the NIM method started with 99.9% of CO2 in the mobile phase because the impurities E-C-B were less polar compared to NIM itself and impurities A-D, and we aimed to retard their elution from the column. In addition, the ABPR parameter of the system was set at 1500 psi to retain them and to obtain a better resolution because of the mobile phase density decreases when the ABPR pressure decreases. 

It may be seen from the presented chromatograms that the methods developed for TH and NIM separate all known impurities and formulation excipients. Both methods were validated with reference to ICH regulations and pharmacopoeial requirements. Validation results were summarised in the Table 2.

From the validation results, the method was analyst friendly, accurate, precise, selective and reproducible. In addition to those, UPC2 was a ‘life-saver’ technique for our laboratory when dealing with this complex product.

 

References

[1] US2011005387A1, F.Cakir, A. Turkyilmaz, L. Oner, U. Cifter, Sanovel Pharmaceuticals, (2011) http://worldwide.espacenet.com/publicationDetails/biblio?CC=US&NR=2011053877A1&KC=A1&FT=D

[2] Suraj D. Jadhav, S.R. Butle, Sachin D. Patil, P.K. Jagtap, Arabian Journal of Chemistry (2011) doi:10.1016/j.arabjc.2011.01.018

[3] Christopher J. Hudalla, Patrick D. McDonald, Waters Corporation, Milford, MA www.waters.com/upc2

Free to read

This article has been unlocked and is ready to read.

Download


Digital Edition

Chromatography Today - Buyers' Guide 2022

October 2023

In This Edition Modern & Practical Applications - Accelerating ADC Development with Mass Spectrometry - Implementing High-Resolution Ion Mobility into Peptide Mapping Workflows Chromatogr...

View all digital editions

Events

Korea Lab 2024

Apr 23 2024 Kintex, South Korea

Korea Chem 2024

Apr 23 2024 Seoul, South Korea

AOCS Annual Meeting & Expo

Apr 28 2024 Montreal, Quebec, Canada

SETAC Europe

May 05 2024 Seville, Spain

ChemUK 2024

May 15 2024 Birmingham, UK

View all events